Case Report
Openaccess
Percutaneous Management of Non Malignant
Superior Vena Cava Syndrome Using IN.PACT
Paclitaxel Coated Balloon
Omer Iftikhar1, Eileen Gajo1, Westby Fisher1, Jonathan Gilbert1 and Michael H. Salinger1*
1Department of Cardiology, NorthShore University Health System, USA
*Corresponding author: Michael H. Salinger, MD, Department of Cardiology, NorthShore University Health System, USA, Tel no: 847-570-2450,
Fax no: 847-570-1865; E-mail:
@
Received: 14 March, 2017 ; Accepted: 12 May, 2017; Published: 22 May, 2017
Citation: Michael H. Salinger, et.al. (2017) Percutaneous Management of Non Malignant Superior Vena Cava Syndrome Using IN.PACT Paclitaxel Coated Balloon. J Cardiovascular Thoracic Surgery 2(3):1-4. DOI: 10.15226/2573-864X/2/3/00117
Abstract
Superior Vena Cava (SVC) syndrome is an uncommon but known
complication of Implantable Cardioverter Defibrillator (ICD) and
pacemaker lead placement. In general 0.2 to 3% of all implants are
associated with symptomatic SVC stenosis. Stent placement is a viable
option for benign SVC syndrome in the absence of indwelling pacing
leads. The risk of perforation and lead fracture increases when large
numbers of leads are present chronically. We describe here a case of
patient with SVC syndrome secondary to chronic ICD and pacemaker
lead placement and a novel use of IN.PACTtm paclitaxel drug coated
balloon to treat the SVC obstruction with a favorable clinical result.
Keywords: Interventional cardiology; Balloon venoplasty; Drug coated balloon
Keywords: Interventional cardiology; Balloon venoplasty; Drug coated balloon
Introduction
Superior Vena Cava (SVC) syndrome occurs when there
is external or internal obstruction of the SVC prior to its entry into
the Right Atrium (RA) [1]. Before the discovery of antibiotics, most
of the cases of SVC syndrome were from infections, particularly
syphilitic aortic aneurysm and tuberculosis [2]. The syndrome is
now more often associated with malignancies, most commonly
lung cancer and Non-Hodgkin Lymphoma (NHL) [3].
With increasing use of pacemakers and defibrillators, the benign causes of SVC syndrome are increasing. Benign causes of SVC syndrome account for about 20-30% cases of the disease [4]. In general 0.2 to 3% of all implants including pacemakers, Implantable Cardioverter Defibrillator (ICD) and Biventricular pacemakers are associated with symptomatic SVC stenosis [5]. Silent venous thrombosis, without any clinical symptoms, is found on routine imaging in as many as 30% of all the implants [6]. We describe here a case of patient with markedly symptomatic SVC syndrome secondary to chronic ICD and pacemaker lead placement and a novel use of IN.PACTtm paclitaxel-coated balloon to dilate the SVC obstruction.
With increasing use of pacemakers and defibrillators, the benign causes of SVC syndrome are increasing. Benign causes of SVC syndrome account for about 20-30% cases of the disease [4]. In general 0.2 to 3% of all implants including pacemakers, Implantable Cardioverter Defibrillator (ICD) and Biventricular pacemakers are associated with symptomatic SVC stenosis [5]. Silent venous thrombosis, without any clinical symptoms, is found on routine imaging in as many as 30% of all the implants [6]. We describe here a case of patient with markedly symptomatic SVC syndrome secondary to chronic ICD and pacemaker lead placement and a novel use of IN.PACTtm paclitaxel-coated balloon to dilate the SVC obstruction.
Case Presentation
74 year old female with a past medical history of
coronary artery disease (previous myocardial infarction at age 31)
sustained ventricular tachycardia with cardiac arrest requiring
ICD placement, paroxysmal atrial fibrillation, remote pulmonary
embolism and ischemic cardiomyopathy with Ejection Fraction
(EF) of 13% presented with complaints of facial swelling,
periorbital edema, and right arm swelling. About 5 months prior,
the patient had her ICD upgraded to a biventricular ICD given her
significantly reduced EF and New York Heart Association (NYHA)
class IV symptoms of heart failure. At the time, the decision was
made to forego removal of her previous ICD wires given that she
had the leads in for a significant amount of time and removal
would be a complex procedure with significant risk of vessel
or cardiac perforation. No stenosis or thrombosis of the SVC or
subclavian veins was found on the venogram done at the time of
the previous device upgrade. The patient tolerated the pacemaker
upgrade procedure well and she was discharged to home the
same day. Months later, the patient started having complaints of
facial swelling, head pressure and upper extremity swelling with
intermittent throat tightness. She initially saw her primary care
physician and was given steroids and her sacubitril/valsartan
Entresto®) was stopped as the symptoms were thought to be
a reaction to the drug. After several weeks, discontinuation of
Entresto and allergy treatments did not alleviate her symptoms.
Other relevant tests that were obtained included chest X-ray that
showed mild cardiomegaly and presence of pacemaker leads in
the SVC (Figure 1).
Figure 1:Chest X-ray (Anteroposterior and lateral view) showing location
of the multiple transvenous leads in the SVC and right atrium
Figure 2:(A) Non contrast coronal section of the CT chest showing
multiple pacemaker leads at the SVC and right atrium and (B) contrast
coronal imaging demonstrating high grade stenosis or occlusion at the
RA/SVC junction with lead artifact making distinction between high
grade stenosis and occlusion difficult
A transthoracic echocardiogram (TTE) showed pacing leads in
the right heart chamber without any evidence of thrombus in
the right atrium. Computed tomography (CT) of the chest was
also obtained which demonstrated the multiple leads and a high
grade narrowing or obstruction of the SVC as it entered the right
atrium (Figure 2a and 2b). Collateral flow was demonstrated into
the azygous and hemiazygous systems of veins. Ultimately, lead
extraction and stenting were felt to represent excessive risk due
to her limited cardiac reserve. The decision was made to proceed
with attempted balloon dilatation of the SVC using a paclitaxel
coated balloon.
The Procedure
Needle entry of the right internal jugular (IJ) vein
with a 5 French introducer was accomplished without difficulty
using ultrasound guidance and micropunture technique. Right IJ
venous pressures were recorded and selective SVC venography
performed. Initial SVC pressure was 26 mmHg. The SVC was
shown to be 100% occluded just below the level of the left
subclavian vein. The left subclavian vein was patent with
very large intrathoracic collaterals presumed to be superior
intercostal/hemiazygous veins (Figure 3a). After review of the
venography, percutaneous transluminal venoplasty (PTV) of the
occluded SVC was performed. The obstruction was crossed using
a 0.035-angled hydrophilic guide wire via a 4F multipurpose
catheter. The 4F catheter was then advanced to the RA and after
confirming appropriate intra-atrial position angiographically,
was used to exchange for a 0.035 extra support guide wire. The
4F system was exchanged over the extra support guide wire for
a 23 cm long 7F sheath, which was positioned in the mid RA. A
4mm x 6 cm Mustang balloon was then advanced via the sheath
and guide wire. After positioning the sheath more proximally the
area of obstruction in the SVC was dilated with the 4 mm balloon
to a maximal pressure of 6 atm with resolution of deformity on
the balloon. Because of tapering present at the tip of sheath, the
technique of advancing the sheath beyond the obstruction to
deliver the balloon was chosen in an attempt to minimize risk of
distal embolization. A 7 mm X 80 cm IN.PACT Paclitaxal eluting
balloon was then positioned across the lesion and inflated at 8
atm pressure (Figure 3b). A 10 x 4 Opti Pro balloon was then used
to perform serial overlapping 5 and 10 min inflations at 4 atm
attempting to modify the potential for elastic recoil and relying
on the effects of the drug coated balloon to reduce post dilatation
intimal fibrosis. Post dilatation, SVC pressure was reduced to
13 mmHg and the stenosis was reduced to 50% with brisk
anterograde flow (Figure 3c). The procedure was well tolerated
and there were no complications or evidence clinically of distal
embolization. The right IJ sheath was removed and pressure
applied to achieve hemostasis. Estimated blood loss was less than
50 ml.
Figure 3a:Venography showing occluded SVC and venous drainage
through the hemi-azygous system of veins
Figure 3b:Percutaneous transluminal venoplasty of the occluded SVC
using IN.PACT Paclitaxal eluting balloon
Figure 3c:Venography post PTV demonstrated restored patency with a
moderate residual stenosis and normal anterograde flow
The patient was admitted for monitoring after her
procedure. She was given a clopidogrel load of 600 mg prior to
the procedure. Heparin bridging was used until she reached a
goal therapeutic INR range between 2.0-3.0 on Coumadin. Total
length of stay in the hospital was 3 days. As per the manufacturer’s
recommendation the patient was put on dual antiplatelet therapy
with aspirin and clopidogrel for one month. Coumadin was
chosen as anti-coagulation for her atrial fibrillation because
of possibility of reversal in case of uncontrolled bleeding. In
addition, after res-establishing antegrade flow venography
appeared to demonstrated residual thrombus at the treatment
site. After 1 month, clopidogrel and Coumadin were discontinued,
aspirin was continued and rivaroxaban was started, as was the
sacubitril/valsartan. She has been followed clinically and has had
no complaints or neck or arm swelling and unchanged normal
pacemaker function now 8 months post therapy
Discussion
Permanent cardiac pacing and ICD placements are
relatively simple and benign procedures that have results in
thousands of lives being saved and improvement of symptoms
in a large number of patients. The procedure is however
associated with a number of complications including infections,
pneumothorax and thrombosis. SVC syndrome is a known
complication of the procedure that occurs due to fibrosis caused
by the pacemaker leads. The greater the number of leads, the
greater the chance of having significant fibrosis and subsequent
SVC obstruction [7].
Treatment options for SVC syndrome include steroids, radiation therapy, chemotherapy, anticoagulants, stenting and surgical repair depending on the etiology of the obstruction. Interventional techniques preferred for malignancy associated SVC syndrome includes stenting whereas open surgical repair is reserved for patients with recurrent severe symptoms. Overall recurrence rates with SVC stents range from 0 to 40%; but in most cases, patency is restored with re-intervention [8]. In case of pacemaker leads associated SVC syndrome, stent placement and balloon venoplasty are viable options. The limitation of traditional balloon venoplasty is a high recurrence rate [9]. Our institution has had limited but favorable outcomes with the use of Drug Coated Balloon (DCB) therapy in juxtacardiac lesions such as recurrent pulmonary venous stenosis [10]. Based on our past experience and risk of lead fracture from stent placement over the indwelling leads and the risks of attempted chronic multiple lead extraction, the decision was made to approach treatment in this patient using the IN.PACTtm Admiral Drug-Coated Balloon (DCB). Our patient presented with facial swelling and peri-orbital edema while on rixaroxaban for her atrial fibrillation which was considered a treatment failure for medical therapy, hence the decision was made to pursue DCB balloon therapy versus conservative management with anticoagulation. This is a primary endovascular therapy whose main indication is for angioplasty after pre-dilatation of de novo or restenotic lesions up to 180 mm in length in superficial femoral or popliteal arterial disease. The paclitaxel coating on the balloon causes inhibition of neo-intimal growth through stabilization of microtubules by preventing their depolymerization during final G2/M phases of cell division. The overall effect increases blood flow and reduces thickening of the artery wall by delivering the drug to the vessel wall. Favorable outcomes have been seen in SFA lesions treated with IN.PACT balloon vs conventional PTCA. In the IN.PACT SFA trial’s 24-month results, DCB showed significantly higher primary patency when compared with PTA (78.9% vs 50.1%; p < 0.001). The rates of clinically directed target vessel revascularization (CD-TLR) were 9.1% and 28.3% (p < 0.001) for the DCB and PTA groups, respectively. The rate of vessel thrombosis was low (1.5% DCB vs. 3.8% PTA; p = 0.243), with no new events reported between 1 and 2 years [11]. The use of drug-coated balloons on the venous circulation has also been studied. The smooth muscle cells in the veins are more responsive to effects of anti-proliferative agents as compared to their arterial counterparts [12]. A recent randomized controlled trial showed improved patency of failing hemodialysis vascular access when paclitaxel coated balloon angioplasty was used for vascular stenosis with cumulative target lesion primary patency of 70% with drug coated balloon versus 25% with conventional balloon angioplasty (HR 0.30, 95% CI 0.12 to 0.71, p < 0.006) [13]. Massmann et al. retrospectively reviewed 27 patients with hemodialysis fistula associated central venous stenosis and found that paclitaxel coated balloon angioplasty showed statistically significant longer freedom period from target vessel revascularization compared to conventional balloon angioplasty [14].
To our knowledge this is the first case of benign SVC syndrome secondary to chronic lead placement that has been treated with IN.PACT drug coated balloon. To date, now 8 months post procedure, the patient remains free of any reoccurrence of signs and symptoms of SVC syndrome and has continued normal pacemaker lead function. Long term follow-up and further studies are needed to evaluate the role of IN.PACT balloon in the treatment of benign SVC syndrome secondary to lead placement.
Treatment options for SVC syndrome include steroids, radiation therapy, chemotherapy, anticoagulants, stenting and surgical repair depending on the etiology of the obstruction. Interventional techniques preferred for malignancy associated SVC syndrome includes stenting whereas open surgical repair is reserved for patients with recurrent severe symptoms. Overall recurrence rates with SVC stents range from 0 to 40%; but in most cases, patency is restored with re-intervention [8]. In case of pacemaker leads associated SVC syndrome, stent placement and balloon venoplasty are viable options. The limitation of traditional balloon venoplasty is a high recurrence rate [9]. Our institution has had limited but favorable outcomes with the use of Drug Coated Balloon (DCB) therapy in juxtacardiac lesions such as recurrent pulmonary venous stenosis [10]. Based on our past experience and risk of lead fracture from stent placement over the indwelling leads and the risks of attempted chronic multiple lead extraction, the decision was made to approach treatment in this patient using the IN.PACTtm Admiral Drug-Coated Balloon (DCB). Our patient presented with facial swelling and peri-orbital edema while on rixaroxaban for her atrial fibrillation which was considered a treatment failure for medical therapy, hence the decision was made to pursue DCB balloon therapy versus conservative management with anticoagulation. This is a primary endovascular therapy whose main indication is for angioplasty after pre-dilatation of de novo or restenotic lesions up to 180 mm in length in superficial femoral or popliteal arterial disease. The paclitaxel coating on the balloon causes inhibition of neo-intimal growth through stabilization of microtubules by preventing their depolymerization during final G2/M phases of cell division. The overall effect increases blood flow and reduces thickening of the artery wall by delivering the drug to the vessel wall. Favorable outcomes have been seen in SFA lesions treated with IN.PACT balloon vs conventional PTCA. In the IN.PACT SFA trial’s 24-month results, DCB showed significantly higher primary patency when compared with PTA (78.9% vs 50.1%; p < 0.001). The rates of clinically directed target vessel revascularization (CD-TLR) were 9.1% and 28.3% (p < 0.001) for the DCB and PTA groups, respectively. The rate of vessel thrombosis was low (1.5% DCB vs. 3.8% PTA; p = 0.243), with no new events reported between 1 and 2 years [11]. The use of drug-coated balloons on the venous circulation has also been studied. The smooth muscle cells in the veins are more responsive to effects of anti-proliferative agents as compared to their arterial counterparts [12]. A recent randomized controlled trial showed improved patency of failing hemodialysis vascular access when paclitaxel coated balloon angioplasty was used for vascular stenosis with cumulative target lesion primary patency of 70% with drug coated balloon versus 25% with conventional balloon angioplasty (HR 0.30, 95% CI 0.12 to 0.71, p < 0.006) [13]. Massmann et al. retrospectively reviewed 27 patients with hemodialysis fistula associated central venous stenosis and found that paclitaxel coated balloon angioplasty showed statistically significant longer freedom period from target vessel revascularization compared to conventional balloon angioplasty [14].
To our knowledge this is the first case of benign SVC syndrome secondary to chronic lead placement that has been treated with IN.PACT drug coated balloon. To date, now 8 months post procedure, the patient remains free of any reoccurrence of signs and symptoms of SVC syndrome and has continued normal pacemaker lead function. Long term follow-up and further studies are needed to evaluate the role of IN.PACT balloon in the treatment of benign SVC syndrome secondary to lead placement.
ReferencesTop
- García Mónaco R, Bertoni H, Pallota G, Lastiri R, Varela M, Beveraggi EM, et al. Use of self-expanding vascular endoprostheses in superior vena cava syndrome. Eur J Cardiothorac Surg. 2003;24(2):208-211.
- Wilson LD, Detterbeck FC, Yahalom J. Superior vena cava syndrome with malignant causes. N Engl J Med. 2007;356(18):1862-1869. doi: 10.1056/NEJMcp067190
- Rice TW, Rodriguez RM, Light RW. The superior vena cava syndrome: clinical characteristics and evolving etiology. Medicine (Baltimore). 2006;85(1):37-42. doi: 10.1097/01.md.0000198474.99876.f0
- Klassen KP, Andrews NC, Curtis GM. Diagnosis and treatment of superior-vena-cava obstruction. AMA Arch Surg. 1951;63(3):311-325.
- Phibbs B, Marriottm HJ. Complications of permanent transvenous pacing. N Engl J Med. 1985;312(22):1428-1432. doi: 10.1056/NEJM198505303122205
- Antonelli D, Turgeman Y, Kaveh Z, Artoul S, Rosenfeld T. Short-term thrombosis after transvenous permanent pacemaker insertion. Pacing Clin Electrophysiol. 1989;12(2):280-282.
- Barakat K1, Robinson NM, Spurrell RA. Transvenous pacing lead-induced thrombosis: a series of cases with a review of the literature. Cardiology. 2000;93(3):142-148.
- Uberoi R. Quality assurance guidelines for superior vena cava stenting in malignant disease. Cardiovasc Intervent Radiol. 2006;29(3):319-322. doi: 10.1007/s00270-005-0284-9
- B. Klop, M G Scheffer, E McFadden, F Bracke, B van Gelder. Treatment of pacemaker-induced superior vena cava syndrome by balloon angioplasty and stenting. Neth Heart J, 2011;19(1):41-46. doi: 10.1007/s12471-010-0052-6
- Rosenberg J, Fisher WG, Guerrero M, Smart S, Levisay J, Feldman T, et al. Drug-Coated Balloon Venoplasty for In-Stent Restenosis in a Patient with Recurrent Pulmonary Vein Stenosis Post Ablation for Atrial Fibrillation: Initial Experience With a New Treatment Technique. J Invasive Cardiol. 2016;28(5):E44-48.
- Laird JR, Schneider PA, Tepe G, Brodmann M, Zeller T, Metzger C, et al., Durability of Treatment Effect Using a Drug-Coated Balloon for Femoropopliteal Lesions: 24-Month Results of IN.PACT SFA. J Am Coll Cardiol. 2015;66(21):2329-2338. doi: 10.1016/j.jacc.2015.09.063
- Kim SJ, Masaki T, Leypoldt JK, Kamerath CD, Mohammad SF, Cheung AK. Arterial and venous smooth-muscle cells differ in their responses to antiproliferative drugs. J Lab Clin Med. 2004;144(3):156-162. doi: 10.1016/j.lab.2004.06.002
- Katsanos K, Karnabatidis D, Kitrou P, Spiliopoulos S, Christeas N, Siablis D. Paclitaxel-coated balloon angioplasty vs. plain balloon dilation for the treatment of failing dialysis access: 6-month interim results from a prospective randomized controlled trial. J Endovasc Ther. 2012;19(2):263-272. doi: 10.1583/11-3690.1
- Massmann A, Fries P, Obst-Gleditsch K, Minko P, Shayesteh-Kheslat R, Buecker A. Paclitaxel-coated balloon angioplasty for symptomatic central vein restenosis in patients with hemodialysis fistulas. J Endovasc Ther. 2015;22(1):74-79. doi: 10.1177/1526602814566907







