2IHU LIRYC, University of Bordeaux – Inserm, Pessac, France
Methods: We designed a randomized comparison between circular versus conventional PVI with the primary end point of reconnection: acutely, in the index procedure, and at 3 months, at the time of a systematic repeat procedure.
Results: Eighty-one patients randomized to circular (nMARQ n = 38) versus conventional (Thermocool Navistar n = 43) underwent PVI for paroxysmal atrial fibrillation. Both groups had similar characteristics: nMARQ age 61 ± 11 years (84% male) and TC age 60 ± 9 years (68% male). NMARQ procedures were shorter (RF and procedure duration), with fewer RIPV reconnections, both acutely and at 3 months (index: 1 (3%) versus 7 (16%), p = 0.044, redo: 21(18%) versus 36 (28%), p = 0.05). Patients ablated with the circular catheter also had shorter repeat procedures suggesting more discrete PV reconnections (nMRAQ: 142 ± 56 versus TC: 186 ± 76 minutes, p = 0.001). At 12 months follow up, 77% of patients in nMARQ group and 79% in TC group were in SR (p = 0.852). No major procedural complications were observed.
Conclusion: This randomized study shows that nMARQ catheter is 3 times faster in isolating the PVs, with shorter RF and procedure time. It is also associated with fewer PV reconnections at 3 months. NMARQ appeared to be safe with no major procedural complications.
Keywords: Paroxysmal atrial fibrillation; Pulmonary vein isolation; Circular (nMARQ) catheter ablation; Single point-by-point (Thermocool catheter ablation); Reconnection rate
However, Atrial Fibrillation (AF) recurrence after PVI is common (20-55%) and generates significant extra costs, including repeat hospital admissions and repeat procedures. PV electrical reconnection is considered the dominant mechanism of recurrence that limits long-term success and remains a current issue despite advances in catheter design, ablation techniques and AF treatment strategies [2-7]. Other potential mechanisms of late recurrence have been described, due to electrical and structural atrial remodelling as a result of aging, heart failure, inflammation, diabetes and obesity [3,8,9].
Recent catheter technologies such as catheter localization, catheter stability and contact force monitoring might be associated with better and more durable Radio Frequency (RF) lesions at first attempt.
Point-by-point standard ablation using single-tip (conventional) catheter is the main current technique for PVI but it can be challenging, complex and time-consuming. A circular, multi-electrode, mapping and ablation catheter (nMARQ) was developed, which provides fast and effective PVI with comparable success and safety [10]. To the best of our knowledge, the circular ablation catheter has not been compared to the conventional point-by-point approach in a randomized study of PVI. Moreover, the acute and late reconnection rate using circular catheter versus point-by-point ablation remains unknown. We designed such a randomized comparison with the primary end point of reconnection rate: acutely, in an index procedure, and at 3 months, in a repeat procedure.
Antiarrhythmic Drugs (AADs), excepting amiodarone, were discontinued for ≥ 5 half-lives before the procedure. Conscious sedation was achieved with intravenous midazolam (0.2 mg/kg), morphine sulphate (0.1 mg/kg to 0.2 mg/kg) and, if necessary, sufentanil (by 5μg up to 20 μg) with continuous noninvasive vital sign monitoring. Surface Electrocardiograms (ECG) and bipolar endocardial ECG were continuously displayed and recorded at a sweep speed of 100 mm/s and filtered from 0.05 to 100 Hz and 30 to 250 Hz, respectively (LabsystemProTM EP recording system, Boston Scientific BSCI, MA, USA).
Using bupivacaine hydrochloride local anaesthesia (100 mg/20 ml subcutaneous SC), three femoral vein sheaths were inserted (8F, 7F and 6F) for vascular access. A steerable quadripolar catheter (Xtrem Sorin, France or Dynamic, Boston scientific, Marlborough, MA, USA) was positioned inside the Coronary Sinus (CS) through the 6F sheath. A single Transeptal Puncture (TSP) was obtained using a 98 cm BRK needle (St. Jude Medical) inserted through a long sheath (Agilis, St. Jude Medical) for patients randomized to nMARQ. For those randomized to point-by-point ablation, the transeptal puncture was carried out using a 78 cm BRK needle and a SL0 long sheath (St. Jude Medical). Access to the LA was confirmed by pressure monitoring and contrast injection (Lobitridol 300 mg l/mL) under fluoroscopy. The long sheath was continuously perfused with heparinized saline (2500 U for 500 ml) at 200 mL/h to reduce the risk of thrombus formation. Immediately after TSP and guide wire insertion, a bolus of intravenous heparin (0.5-0.8 mg/kg) was injected. Heparin was monitored throughout the procedure every 20 to 30 minutes by Activated Clotting Time (ACT) with a target of 300-400 seconds.
According to treatment group, either a circular (nMARQ®, Biosense Webster) or a conventional (NAVISTAR® THERMOCOOL®, Biosense Webster) catheter was used for mapping and ablation. When the nMARQ catheter was used, it was inserted through the long sheath.
In the conventional group, a mapping circular catheter (Lasso®, Biosense Webster) catheter was introduced through the long sheath (described below).
The nMARQ catheter has 10 open irrigated electrodes used for both mapping and ablation (electrode length 3.5 mm, spacing 4 mm, maximum catheter diameter 8.4 F) arranged on an adaptable circle (diameter 20-35 mm), to allow optimal positioning in the PV antrum (Figure 1 and 2) [10]. The catheter was irrigated at 60 mL/min 0.9 saline during RF delivery (Cool Flow, Biosense Webster). The generator (nMARQ Generator, Biosense Webster) delivers RF energy through the 10 separated electrodes in unior bipolar mode. Maximum power used was 25 Watt (W) (20 W for posterior wall to reduce the risk of oesophageal injury), temperature 45°C, maximum duration time 60 seconds in unipolar mode. Catheter position, placement and contact was monitored and optimized before any RF application. Contact was assessed visually with fluoroscopy and also by impedance-based technology provided by Carto® 3 (Tissue Connect, Biosense Webster). RF parameters were continuously monitored during ablation and intermittent fluoroscopy was obtained to assess diaphragm movement, in order to prevent right phrenic nerve injury during right PV ablation. For confirmation of PV isolation, the circular catheter was moved slightly more distally before and after ablation (using the smallest diameter) to verify far-field signals and assess for entrance block. Pacing for phrenic nerve capture and oesophageal temperature monitoring were not routinely used.
ECG: Surface Electrocardiogram; PV: Pulmonary Vein; RSPV: Right Superior Pulmonary Vein; SR: Sinus Rhythm; CS: Coronary Sinus; AP: AnteroPosterior
AP: Antero-Posterior; PA: Postero-Anterior
AP: Anteroposterior; PA: Posteroanterior
ECG: Surface Electrocardiogram; PV: Pulmonary Vein; LSPV: Left Superior Pulmonary Vein; LIPV: Left Inferior Pulmonary Vein; SR: Sinus Rhythm; AP: Anteroposterior
Cavotricuspid Isthmus (CTI) ablation was performed for patients with documented typical atrial flutter in both groups. LA defragmentation ± Direct Current cardioversion (DC) shock/ AADs were allowed if AF persisted after PVI. The end point of LA defragmentation was to reduce signal complexity and to prolong the local cycle length, to finally achieve AF termination to SR or atrial tachycardia.
The procedure was performed as for the index procedure with conventional catheters. LA and PV 3D anatomy were reconstructed and high-density anatomical mapping was acquired with Carto 3 system using TC and/or Lasso NAV. Each PV antral region was carefully evaluated for isolation or electrical reconnections/gaps in PV line using Lasso mapping catheter. PVs were divided into 4 segments (superior, anterior, inferior and posterior) and each segment with reconnection was targeted with RF ablation (TC) until complete PVI was achieved (Figure 5). PVI was confirmed locally by the Lasso catheter, with pacing for entrance and exit block and adenosine as explained in index procedure.
The CTI line if present was rechecked for bidirectional block and re-ablated again if necessary. Additional LA defragmentation
AP: Anteroposterior; PA: Posteroanterior
N=81 |
nMARQ(n 38) |
Navistar TC (n 43) |
p |
Male Sex, n (%) |
32(84) |
29(68) |
0.081 |
Age, mean ± SD (years) |
61 ± 11 |
60 ± 9 |
0.613 |
BMI*, mean ± SD (kg/m2) |
25.5 ± 3.8 |
26.8 ± 3.8 |
0.118 |
Hypertension, n (%) |
8 (21) |
13 (30) |
0.347 |
Diabetes Mellitus, n (%) |
1 (3) |
1 (2) |
0.929 |
Embolic events, n (%) |
3 (8) |
5 (12) |
0.574 |
Structural heart disease, n (%) |
|
|
0.555 |
Ischemic heart disease |
4 (11) |
1 (2) |
0.126 |
Dilated or hypertrophic cardiomyopathy |
2 (5) |
1 (2) |
0.484 |
Valvular heart disease |
5 (13) |
6 (14) |
0.917 |
Echocardiographic parameters |
|||
Left ventricular ejection fraction, mean ± SD (%) |
63 ± 6 |
61 ±6 |
0.231 |
Left atrial surface area, mean ± SD (cm2) |
18.9 ± 5.0 |
19.5 ± 5.0 |
0.607 |
AF† related parameters |
|||
AF history, mean ± SD (months) |
71 ± 77 |
89 ± 86 |
0.352 |
AF max duration episodes, mean ± SD (hours) |
23.6 ± 26.9 |
27.3 ± 38.7 |
0.657 |
AF at baseline, n (%) |
8 (21) |
8 (19) |
0.782 |
≥ 1 DC* shock, n (%) |
4 (10) |
6 (14) |
0.743 |
AADs† used before AF ablation, n (%) |
36 (95) |
41 (95) |
1.0 |
Amiodarone use before AF ablation, n (%) |
18 (47) |
16 (37) |
0.485 |
CHA2DS2-VASc, mean ± SD |
1 ± 1 |
1 ± 1 |
0.388 |
†Atrial fibrillation
*DC, direct current cardio version
†AAD, antiarrhythmic drug
min ± SD |
nMARQ (n 38) |
Navistar TC (n 43) |
p |
RSPV† |
4 ± 3 |
11 ± 8 |
< 0.001 |
RIPV‡ |
3 ± 2 |
11 ± 5 |
< 0.001 |
LSPV§ |
3 ± 2 |
12 ± 6 |
< 0.001 |
LIPV** |
2 ± 2 |
8 ± 5 |
< 0.001 |
RPV†† |
6 ± 4 |
20 ± 9 |
< 0.001 |
LPV‡‡ |
5 ± 3 |
18 ± 8 |
< 0.001 |
Total PV |
12 ± 5 |
38 ± 13 |
< 0.001 |
†Right superior pulmonary vein
‡Right inferior pulmonary vein
§Left superior pulmonary vein
**Left inferior pulmonary vein
††Right pulmonary veins
‡‡Left pulmonary veins
A lower reconnection rate in RIPV in nMARQ group was observed, compared to TC, 3% versus 16%, p = 0.044). RSPV in nMARQ group and LSPV in TC group had a higher reconnection rate when compared to other veins of the same catheter group (Figure 6).
The acute reconnection rate per PV segment was available in 11 (73%) patients in the nMARQ arm and 17 (74%) patients in the TC arm as shown in table 1 in supplementary files. The reconnection rate in the anterior segment of RSPV was significantly lower in TC compared to nMARQ group 12% versus 46% respectively, p = 0.044). The latter had the highest reconnection rate compared to other PV segments in nMARQ group. On the other hand, the inferior segment of RIPV and the anterior segment of LSPV (24%) had the highest reconnection rate in TC group.
There was no significant difference in the number of acute reconnection per patients between both groups (Figure 7). Seventy percent of patients had no acute reconnection in nMARQ versus 58% in TC group. In some cases, PV reconnection was limited to one vein (22% with nMARQ versus 30% with TC). The total number of acutely reconnected PV was higher in the TC group but this was not statistically significant (p = 0.351).
There was no significant statistical difference between both groups in non-PV ablations and DC/AADs. RF duration for non- PV ablation was significantly lower in nMARQ group compared to TC (5 ± 3 versus 15 ± 11 min, p = 0.014), respectively shown in table 3 in the supplementary file.
n (%) |
nMARQ (n 11) |
Navistar TC (n 17) |
p |
|
RSPV |
Superior |
0 (0) |
0 (0) |
- |
Anterior |
5 (46) |
2 (12) |
0.044 |
|
Inferior |
1 (9) |
2 (12) |
0.823 |
|
Posterior |
1 (9) |
1 (6) |
0.747 |
|
RIPV |
Superior |
0 (0) |
3 (18) |
0.14 |
Anterior |
0 (0) |
3 (18) |
0.14 |
|
Inferior |
0 (0) |
4 (24) |
0.082 |
|
Posterior |
1 (9) |
2 (12) |
0.823 |
|
LSPV |
Superior |
1 (9) |
2 (12) |
0.823 |
Anterior |
3 (27) |
4 (24) |
0.823 |
|
Inferior |
1 (9) |
2 (12) |
0.823 |
|
Posterior |
0 (0) |
0 (0) |
- |
|
LIPV |
Superior |
1 (9) |
1 (6) |
0.747 |
Anterior |
1 (9) |
1 (6) |
0.747 |
|
Inferior |
1 (9) |
1 (6) |
0.747 |
|
Posterior |
1 (9) |
1 (6) |
0.747 |
|
nMARQ (n 37) |
Navistar TC (n 43) |
p |
|
LA defragmentation, n (%) |
8 (22) |
5 (12) |
0.226 |
CTI, n (%) |
10 (27) |
6 (14) |
0.145 |
RF duration, mean (min) ± SD |
5 ± 3 |
15 ± 11 |
0.014 |
DC, n (%) |
2 (5) |
8 (19) |
0.075 |
Flecainide, n (%) |
2 (5) |
6 (14) |
0.203 |
n (%) |
nMARQ (n 38 ) |
Navistar TC (n 43) |
p |
On AADS @ 12 months |
18 (45) |
20 (44) |
ns* |
Flecainide |
4 (10) |
5 (11) |
ns |
Amiodarone |
1 (3) |
1 (2) |
ns |
Beta blockers |
16 (40) |
17 (38) |
ns |
Ca2+ channel blockers |
1 (3) |
2 (4) |
ns |
Propafenone |
0 (0) |
1 (2) |
ns |
nMARQ (n 30) |
Navistar TC (n 35) |
p |
|
LA defragmentation, n (%) |
5 (17) |
11 (31) |
0.168 |
CTI, n (%) |
8 (27) |
12 (34) |
0.507 |
RF duration, mean (min) ± SD |
22 ± 14 |
20 ±16 |
0.818 |
DC, n (%) |
1 (3) |
2 (6) |
0.648 |
There was no link between PV reconnection per vein or per segment between index and repeated procedures (Cohen’s kappa = 0.048 and – 0.042).
Two patients (3%) had puncture-related complications after repeat procedures, including 1 arteriovenous fistula and 1 femoral artery aneurysm. One (1%) patient had an acute pericardial effusion in nMARQ group during the index procedure. He was percutaneously drained with no consequences. There were no incidences of stroke, phrenic nerve paralysis or esophageal fistula.
2. In index procedures, additional ablation at non-PV sites (when needed) was faster when ablated with nMARQ catheter.
3 months n (%) |
nMARQ (n 37) |
Navistar TC (n 43) |
p |
Symptoms |
17 (46) |
16 (37) |
0.429 |
Holter + |
18 (49) |
15 (35) |
0.212 |
6 months |
nMARQ (n 32) |
Navistar TC (n 39) |
p |
Symptoms |
5 (16) |
9 (23) |
0.432 |
Holter + |
5 (16) |
6 (15) |
0.978 |
12 months |
nMARQ (n 35) |
Navistar TC (n 38) |
P |
Symptoms |
10 (29) |
12 (32) |
0.78 |
Holter + |
8 (23) |
8 (21) |
0.852 |
min ± SD |
nMARQ (n 25) |
Navistar TC (n 28) |
p |
RSPV |
7 ± 5 |
6 ± 5 |
0.744 |
RIPV |
7 ± 5 |
6 ± 5 |
0.843 |
LSPV |
7 ± 7 |
9 ± 5 |
0.619 |
LIPV |
5 ± 6 |
5 ± 4 |
0.791 |
RPV |
8 ± 5 |
10 ± 6 |
0.239 |
LPV |
7 ± 8 |
10 ± 7 |
0.308 |
n (%) |
nMARQ (n 30) |
Navistar TC (n 32) |
p |
|
RSPV |
Superior |
5 (17) |
5 (17) |
1 |
Anterior |
3 (10) |
6 (20) |
0.279 |
|
Inferior |
5 (17) |
3 (10) |
0.448 |
|
Posterior |
4 (13) |
9 (30) |
0.117 |
|
RIPV |
Superior |
3 (10) |
3 (9) |
0.933 |
Anterior |
6 (20) |
9 (28) |
0.455 |
|
Inferior |
8 (27) |
13 (41) |
0.246 |
|
Posterior |
4 (13) |
11 (34) |
0.053 |
|
LSPV |
Superior |
7 (23) |
6 (19) |
0.658 |
Anterior |
3 (10) |
13 (41) |
0.006 |
|
Inferior |
3 (10) |
2 (6) |
0.588 |
|
Posterior |
6 (20) |
6 (19) |
0.901 |
|
LIPV |
Superior |
4 (13) |
3 (9) |
0.622 |
Anterior |
2 (7) |
3 (9) |
0.696 |
|
Inferior |
5 (17) |
4 (13) |
0.642 |
|
Posterior |
4 (13) |
5 (16) |
0.798 |
|
n/16 (%) |
nMARQ (n 30) |
Navistar TC (n 32) |
p |
0 |
5 (17) |
4 (13) |
0.728 |
1 |
5 (17) |
3 (9) |
0.467 |
2 |
8 (27) |
8 (25) |
1 |
3 |
5 (17) |
7 (22) |
0.751 |
4 |
4 (13) |
3 (9) |
0.703 |
5 |
- |
1 (3) |
1 |
6 |
1 (3) |
2 (6) |
1 |
7 |
2 (7) |
1 (3) |
0.607 |
8 |
- |
2 (6) |
0.492 |
9 |
- |
1 (3) |
1 |
Total PV segment reconnection |
25 (83) |
27 (87) |
0.728 |
Total PV segment reconnection extent, n ± SD |
2 ± 2 |
3 ± 2 |
0.177 |
4. RIPV reconnection was less common after ablation with the nMARQ catheter, both acutely and at 3 months.
5. Patients ablated with the circular catheter also had shorter repeat procedures suggesting more discrete PV reconnections.
6. The circular ablation catheter appeared to be safe with no major procedural complications.
The RF duration to isolate the veins has been reported to vary from 10 to 14 minutes , with procedure time from 86 to 131 minutes and X ray exposure from 4 to 26 minutes [10-12,15,16].
Acute reconnection has been reported in 22% (Lasso only) to 42% when adenosine was used [10,12]. Our findings confirm that, when using the nMARQ catheter, the RSPV is the most challenging vein, requiring more RF applications and with a higher acute reconnection rate per vein and per segment; particularly at the anterior segment. On the other hand, RIPV showed a significantly lower reconnection rate per vein in nMARQ acutely and at 3 months, particularly at the posterior segment when compared to TC.
Our study shows heterogeneous efficacy of the nMARQ catheter, with superiority on the RIPV, particularly on the posterior segment, and at the ridge between LAA and LSPV. This may reflect better catheter stability and contact. This observation may guide improvements in performance on other veins and segments.
In a large, multicentre study, 374 patients underwent PVI with nMARQ for PAF and persistent AF (PsAF) in 5 centers with an acute PVI rate of 99.6%. Again, RF times were longer in the superior veins compared to the inferior veins (LSPV 4 ± 3, RSPV 4 ± 2, LIPV and RIPV 3 ± 2 minutes, p = 0.0001). And the anterior aspect of RSPV lesion was also identified as a weak point using this catheter.
PV reconduction is the dominant finding at repeat procedures for AF recurrence, and has been observed in 94% of patients after point-by-point ablation; which may reflect areas of poor catheter contact and lesion discontinuity [3].
Different “one shot” catheter designs and technologies aiming at reducing procedure and fluoroscopy time for PVI have been considered and tested over the last decade. In our study, procedure time to achieve PVI with nMARQ was 111 ± 35 minutes, with fluoroscopy of 26 ± 11 minutes. In a meta-analysis of 23 cryoablation studies, the mean procedure time was 206 ± 72 minutes, with fluoroscopy of 46 ± 13 minutes [17]. In another meta-analysis of 42 studies of AF ablation with PVAC catheter, average procedure time were reported as 117 ± 33 minutes, with fluoroscopy of 27 ± 10 minutes [18].
Serious safety concerns regarding Silent Cerebral Lesions (SCL) caused by multi-electrode ablation in the LA have arisen recently [22,23]. Scaglione reported no major procedural complications and no SCL by 24 hours post procedural cerebral diffusion-weighted DW-MRI, in contrast with the results reported by Deneke which showed that 14 out of 43 patients (33%) had SCL following nMARQ ablation [12,13]. We did not routinely acquire cerebral imaging post ablation.
The complication rate in cryoablation reported in a systematic review was relatively low with 6% incidence of Phrenic Nerve Palsy (PNP), most of them being transient [17]. In a systematic review of PVAC ablation, the overall procedural complication rate appeared to be low (acute procedural complications were 2%, though non-uniform screening may limit the accuracy of this estimate). However, high rates of silent cerebral ischemic lesions (38%-45%) have been reported, significantly higher than with irrigated RF (7%-17%) or cryoablation (4%-6%) [18].
Another single operator study reported 83% success rate at 1 year with the nMARQ [16]. In our study, 77 % of patients in nMARQ group and 79% in TC group were in SR (p = 0.852) at 1 year. Long-term success (≈ 18 months) with freedom from AF, AFL or AT was 80% in a study comparing nMARQ to TC Smart Touch in 86 patients with PAF and PsAF with no significant difference in both groups (p = 0.78) [15].
We did not routinely monitor oesophageal temperature during ablation or perform post ablation esophageal endoscopy or cerebral imaging (MRI). However, we did not observe any oesophageal or cerebral complications, possibly as a consequence of reduced power and duration of RF delivery in the posterior LA.
Another potential limitation was in the technology used in conventional single electrode catheters. We used an impedancebased method rather than direct measurements of force sensing technologies ‘contact force’ that associated with a more complete ablation.
Our observation of nMARQ efficacy and safety suggests that this catheter may be a promising alternative to conventional point-by-point AF ablation and the knowledge gained using this catheter should be used to build better ablation tools in the future.
- Fuster V, Ryden LE, Asinger RW, Cannom DS, Crijns HJ, Frye RL, et al. ACC/AHA/ESC Guidelines for the Management of Patients With Atrial Fibrillation: Executive Summary A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines and the European Society of Cardiology Committee for Practice Guidelines and Policy Conferences (Committee to Develop Guidelines for the Management of Patients With Atrial Fibrillation) Developed in Collaboration With the North American Society of Pacing and Electrophysiology. Circulation. 2001. 104:2118-2150.
- Wilber DJ, Pappone C, Neuzil P, De Paola A, Marchlinski F, Natale A, et al. Comparison of antiarrhythmic drug therapy and radiofrequency catheter ablation in patients with paroxysmal atrial fibrillation: a randomized controlled trial. JAMA. 2010;303(4):333-340. doi: 10.1001/jama.2009.2029
- Ouyang F, Tilz R, Chun J, Schmidt B, Wissner E, Zerm T, et al. Long-term results of catheter ablation in paroxysmal atrial fibrillation: lessons from a 5-year follow-up. Circulation. 2010;122(23):2368-2377. doi: 10.1161/CIRCULATIONAHA.110.946806
- Hussein AA, Saliba WI, Martin DO, Bhargava M, Sherman M, Magnelli-Reyes C, et al. Natural history and long-term outcomes of ablated atrial fibrillation. Circ Arrhythm Electrophysiol. 2011;4(3):271-278. doi: 10.1161/CIRCEP.111.962100
- Natale A, Raviele A, Arentz T, Calkins H, Chen SA, Haissaguerre M, et al. Venice Chart international consensus document on atrial fibrillation ablation. J Cardiovasc Electrophysiol. 2007;18(5):560-580.
- Daubert JC, Saxon L, Adamson PB, Auricchio A, Berger RD, Beshai JF, et al. 2012 EHRA/HRS expert consensus statement on cardiac resynchronization therapy in heart failure: implant and follow-up recommendations and management. Europace. 2012;14(9):1236-1286. doi: 10.1093/europace/eus222
- Calkins H, Kuck KH, Cappato R, Brugada J, Camm AJ, Chen SA, et al. 2012 HRS/EHRA/ECAS expert consensus statement on catheter and surgical ablation of atrial fibrillation: recommendations for patient selection, procedural techniques, patient management and follow-up, definitions, endpoints, and research trial design: a report of the Heart Rhythm Society (HRS) Task Force on Catheter and Surgical Ablation of Atrial Fibrillation. Developed in partnership with the European Heart Rhythm Association (EHRA), a registered branch of the European Society of Cardiology (ESC) and the European Cardiac Arrhythmia Society (ECAS); and in collaboration with the American College of Cardiology (ACC), American Heart Association (AHA), the Asia Pacific Heart Rhythm Society (APHRS), and the Society of Thoracic Surgeons (STS). Endorsed by the governing bodies of the American College of Cardiology Foundation, the American Heart Association, the European Cardiac Arrhythmia Society, the European Heart Rhythm Association, the Society of Thoracic Surgeons, the Asia Pacific Heart Rhythm Society, and the Heart Rhythm Society. Heart Rhythm. 2012;9(4):632-696.e21. doi: 10.1016/j.hrthm.2011.12.016
- Weerasooriya R, Khairy P, Litalien J, Macle L, Hocini M, Sacher F, et al. Catheter ablation for atrial fibrillation: are results maintained at 5 years of follow-up? J Am Coll Cardiol. 2011;57(2):160-166. doi: 10.1016/j.jacc.2010.05.061
- Wong CX, Abed HS, Molaee P, Nelson AJ, Brooks AG, Sharma G, et al. Pericardial fat is associated with atrial fibrillation severity and ablation outcome. J Am Coll Cardiol. 2011;57(17):1745-51. doi: 10.1016/j.jacc.2010.11.045
- Zellerhoff S, Daly M, Lim HS, Denis A, Komatsu Y, Jesel L, et al. Pulmonary vein isolation using a circular, open irrigated mapping and ablation catheter (nMARQ): a report on feasibility and efficacy. Europace. 2014;16(9):1296-1303. doi: 10.1093/europace/euu133
- Mahida S, Hooks DA, Nentwich K, Ng GA, Grimaldi M, Shin DI, et al. nMARQ Ablation for Atrial Fibrillation: Results from a Multicenter Study. J Cardiovasc Electrophysiol. 2015;26(7):724-729. doi: 10.1111/jce.12698
- Scaglione M, Caponi D, Anselmino M, Di Clemente F, Blandino A, Ferraris F, et al. Pulmonary vein isolation with a new multipolar irrigated radiofrequency ablation catheter (nMARQ): feasibility, acute and short-term efficacy, safety, and impact on postablation silent cerebral ischemia. J Cardiovasc Electrophysiol. 2014;25(12):1299-1305. doi: 10.1111/jce.12500
- Deneke T, Schade A, Muller P, Schmitt R, Christopoulos G, Krug J, et al. Acute safety and efficacy of a novel multipolar irrigated radiofrequency ablation catheter for pulmonary vein isolation. J Cardiovasc Electrophysiol. 2014 Apr;25(4):339-45. doi: 10.1111/jce.12316
- Rosso R, Chorin E, Levi Y, Rogowski O, Viskin S. Radiofrequency Ablation of Atrial Fibrillation: Nonrandomized Comparison of Circular versus Point-by-Point "Smart" Ablation for Achieving Circumferential Pulmonary Vein Isolation and Curing Arrhythmic Symptoms. J Cardiovasc Electrophysiol. 2016. doi: 10.1111/jce.13058
- Rosso R, Halkin A, Michowitz Y, Belhassen B, Glick A, Viskin S. Radiofrequency ablation of paroxysmal atrial fibrillation with the new irrigated multipolar nMARQ ablation catheter: Verification of intracardiac signals with a second circular mapping catheter. Heart Rhythm. 2014;11(4):559-565. doi: 10.1016/j.hrthm.2013.12.029
- Martin Vurma, Lam Dang, Hans-Peter Brunner-La Rocca, Gabor Sutsch, Christine H Attenhofer-Jost, Firat Duru, et al. Safety and efficacy of the nMARQ catheter for paroxysmal and persistent atrial fibrillation. Europace. 2016;18(8):1164-1169.
- Andrade JG, Khairy P, Guerra PG, Deyell MW, Rivard L, Macle L, et al., Efficacy and safety of cryoballoon ablation for atrial fibrillation: a systematic review of published studies. Heart Rhythm. 2011;8(9):1444-1451. doi: 10.1016/j.hrthm.2011.03.050
- Andrade JG, Dubuc M, Rivard L, Guerra PG, Mondesert B, Macle L, et al. Efficacy and safety of atrial fibrillation ablation with phased radiofrequency energy and multielectrode catheters. Heart Rhythm. 2012;9(2):289-296. doi: 10.1016/j.hrthm.2011.09.009
- Shin DI, Kirmanoglou K, Eickholt C, Schmidt J, Clasen L, Butzbach B, et al. Initial results of using a novel irrigated multielectrode mapping and ablation catheter for pulmonary vein isolation. Heart Rhythm. 2014;11(3):375-383. doi: 10.1016/j.hrthm.2013.12.008
- Rillig A, Lin T, Burchard A, Kamioka M, Heeger C, Makimoto H, et al. Modified energy settings are mandatory to minimize oesophageal injury using the novel multipolar irrigated radiofrequency ablation catheter for pulmonary vein isolation. Europace. 2015;17(3):396-402. doi: 10.1093/europace/euu269
- Cappato R, Calkins H, Chen SA, Davies W, Iesaka Y, Kalman J, et al. Prevalence and causes of fatal outcome in catheter ablation of atrial fibrillation. J Am Coll Cardiol. 2009;53(19):1798-1803. doi: 10.1016/j.jacc.2009.02.022
- Herrera Siklody C, Deneke T, Hocini M, Lehrmann H, Shin DI, Miyazaki S, et al. Incidence of asymptomatic intracranial embolic events after pulmonary vein isolation: comparison of different atrial fibrillation ablation technologies in a multicenter study. J Am Coll Cardiol. 2011;58(7):681-688. doi: 10.1016/j.jacc.2011.04.010
- Stephan Zellerhoff, Martin A Ritter, Simon Kochhauser, Ralf Dittrich, Julia Kobe, Peter Milberg, et al. Modified phased radiofrequency ablation of atrial fibrillation reduces the number of cerebral microembolic signals. Europace. 2014;16(3):341-346.
- Ganesan AN, Shipp NJ, Brooks AG, Kuklik P, Lau DH, Lim HS, et al. Long-term outcomes of catheter ablation of atrial fibrillation: A systematic review and meta analysis. J Am Heart Assoc. 2013;2(2):e004549. doi: 10.1161/JAHA.112.004549
- Packer DL, Kowal RC, Wheelan KR, Irwin JM, Champagne J, Guerra PG, et al. Cryoballoon ablation of pulmonary veins for paroxysmal atrial fibrillation: First results of the North American Arctic Front (STOP AF) pivotal trial. J Am Coll Cardiol. 2013;61(16):1713-1723. doi: 10.1016/j.jacc.2012.11.064
- Bittner A, Monnig G, Zellerhoff S, Pott C, Kobe J, Dechering D, et al. Randomized study comparing duty-cycled bipolar and unipolar radiofrequency with point-by-point ablation in pulmonary vein isolation. Heart Rhythm. 2011;8(9):1383-1390. doi: 10.1016/j.hrthm.2011.03.051